Maintenance of level meters: cleaning of the radar antenna, inspection of guided waveguides and calibration

Technical analysis: Level transmitter maintenance: radar antenna cleaning, guided wave probe inspection, and calibration

Scope and purpose

This practical guide covers routine maintenance of industrial level gauges of the radar type and guided wave radar (guided wave radar) used to measure the level of liquids and bulk materials in tanks, hoppers and process vessels. The procedures are intended for the maintenance of devices in the chemical, oil refining, food and energy industries in accordance with DSTU EN 61508-1:2018 and ISO 55000:2014.

Maintenance is performed every 3-6 months depending on operating conditions, or when the readings deviate more than ±2% from the control values. Critically important for systems with an accuracy class of 0.2% according to DSTU EN 60770-1.

Safety measures

ATTENTION! Before starting work, it is mandatory to turn off the power supply and block energy carriers (LOTO) according to DSTU OHSAS 18001.

Dangerous factors:

  • High voltage 24-230V DC/AC
  • Radio frequency radiation 26 GHz (SAR level < 1.6 W/kg)
  • Chemicals in Tanks - MSDS Check Required
  • Working at a height of more than 1.8 m - the use of a safety belt is mandatory

Necessary tools and materials

NameSpecificationQuantity
Torque wrench5-50 Nm, accuracy ±3%1
Digital multimeterTrue RMS, 4-20mA, 1000V AC/DC1
Current loop calibrator4-20mA, accuracy ±0.02%1
End headsM6, M8, M10 (6-sided)1 set
Horn keys8, 10, 13, 17 mm1 set
Nylon bristle brushDiameter 25 mm, length 150 mm2
Napkins are lint-freeCellulose, size 200x200 mm20
Technical solventIsopropanol 99.9%0.5 l
Silicone greaseTemperature range -40...+200°C50 g
O-ring sealsNBR 70 Shore A (various sizes)kit

Inspection before maintenance

Check elementWhat to checkAcceptance/rejection criteriaNotes
Device bodyMechanical damage, corrosionReject: cracks >0.5 mm, corrosion >5% of the surfacePhotofixation of defects
Cable inputsConnection density, integrityReject: loosening >0.5 revolutionsTightening torque 8-12 Nm
Radar sensor antennaPollution, plaque, deformationReject: plaque >1mm, deformationTeflon coatings are particularly sensitive
Waveguide (GWR)Mechanical damage to the rodReject: bending >2mm, cuts >0.1mmReplacement of the rod in case of damage
Indication of the deviceDeviation from control valuesReject: error >±2% of rangeComparison with the reference level gauge
Output signal current4-20mA according to the levelAccept: linearity ±0.1mAMeasurements on terminals 3-4
Operating temperatureThe surrounding environmentAccept: -40...+80°C for standard executionVerification of compliance of the passport
Mounting vibrationRigidity of installationReject: backlash >0.5mmFlange tightening torque 35-45 Nm

Step-by-step service procedure

Preparatory works

  1. Perform the lockout and tagging (LOTO) procedure: turn off the power at the switchboard, hang a tag with the name of the performer and the date. A common mistake: working under voltage can lead to electric shock up to 230V.
  2. Record the current readings of the device in the service log: value 4-20mA, corresponding level in mm, fixation time.
  3. Photograph the general condition of the equipment for comparison after service.
  4. Check the body surface temperature - should be <60°C for safe touch. At a higher temperature, wait for cooling.

Dismantling and access to nodes

  1. Disconnect the power cable from the terminal box of the level meter. The moment of unscrewing the cable entry: 8-12 Nm (counter-clockwise).
  2. Open the cover of the electronic unit with a 10 mm key. Attention: there may be condensation inside - carefully wipe with a napkin.
  3. For radar devices: unscrew the protective cover of the antenna (torque: 15-20 Nm). Visual indicator of correct dismantling - the antenna is completely open, there are no distortions.
  4. For GWR devices: loosen the clamping nuts of the waveguide (torque: 12-18 Nm). Do not pull the rod completely out of the tank.

Cleaning the radar antenna

  1. Visually assess the type and degree of contamination of the antenna. Typical contaminants: dust (gray coating), product condensate (colored spots), salt crystals (white deposits).
  2. For light dirt: wipe the antenna with a tissue moistened with isopropanol. Movements - from the center to the edges, without strong pressure.
  3. For heavy dirt: use a nylon brush with a diameter of 25 mm. It is forbidden to use metal brushes - damage to the antenna coating!
  4. Check cleanliness: the surface of the antenna should be uniformly matte, without stains and deposits. Unacceptable solvent residues (visually - wet spots).
  5. Dry the antenna in the air for 5-10 minutes. Acceleration of drying with compressed air is allowed at pressure <2 bar.

Inspection of the guided waveguide

  1. Inspect the waveguide rod at the access length (usually 300-500mm from the flange). Look for: bends over 2mm, scratches over 0.1mm deep, product deposits.
  2. Measure the diameter of the rod at three points with a caliper: near the flange, in the middle of the accessible area, at the end. The deviation from the nominal diameter should not exceed ±0.1mm.
  3. Clean the accessible part of the rod with isopropanol and a lint-free cloth. Move only along the rod axis, not in a circle!
  4. Check rod connecting elements (if any): tightening torque 25-35 Nm for M12 connections, 15-25 Nm for M8.
  5. Visually assess the condition of insulators (usually Teflon): absence of cracks, chips, discoloration (yellowing indicates overheating).

Inspection of electronic components

  1. Inspect the boards of the electronic unit: no oxidation of contacts, swelling of capacitors, darkening of resistors.
  2. Measure the insulation resistance between the input terminals and the case with a 500V megohmmeter: it should be >10 MΩ at normal humidity.
  3. Check the integrity of the connecting wires with a multimeter: the resistance of each conductor is <1 Ohm for a length of up to 5 m.
  4. Inspect the terminal connections: the tightening torque of the terminal screws is 0.8-1.2 Nm (use a screwdriver with a torque limiter).

Device calibration

  1. Connect the current loop calibrator to the output terminals (usually 3 and 4). Install a 24V DC power supply.
  2. Put the device in the setting mode according to the manufacturer's instructions (usually a combination of buttons or DIP switches).
  3. Set Empty Tank Value (4mA): Set the reference level to 0% and adjust the output to 4,000mA ±0.02mA.
  4. Set Full Tank Value (20mA): Set the reference level to 100% and adjust the output to 20,000mA ±0.02mA.
  5. Check linearity at intermediate points: 25% (7,000mA), 50% (12,000mA), 75% (16,000mA). The deviation should not exceed ±0.1mA.
  6. Save the calibration in the device's memory and exit the setup mode. Common mistake: Forgetting to save settings results in loss of calibration.

Collection and inspection

  1. Install the protective cover of the antenna or clamp the waveguide according to the specification (points are specified in clause 7-8).
  2. Close the cover of the electronic unit, check the correct fit of the O-ring seal.
  3. Connect the power cable, tighten the cable entry with a torque of 8-12 Nm.
  4. Apply power and wait for the initialization of the device (usually 30-60 seconds, indicated by the blinking of the LED).
  5. Check the current of the output signal at the current level - it should correspond to the calibration of ±0.1mA.

Control check after service

TestExpected resultActual resultPass/Fail
Output signal at minimum level4.000mA ±0.05mA______mA_______
Output signal at maximum level20,000mA ±0.05mA______mA_______
Linearity at the midpoint12,000mA ±0.1mA at 50% level______mA_______
Response time to a change in level<5 seconds for a 10% change______sec_______
Insulation resistance>10 MΩ at 500V______MOm_______
Body temperature in operation<60°C at +25°C ambient______°C_______
No diagnostic errorsGreen light or code 00______________
Mechanical fixation of the caseAbsence of backlash >0.5 mm______________

Diagnostics and troubleshooting

SymptomProbable causeElimination
The output current is blocked at 3.6mAComplete loss of signal, dirty antennaClean the antenna, check the range settings
The output current is blocked at 22mAOut of range, electronics errorCheck installation height, recalibrate
Unstable readings (fluctuation ±0.5mA)Foam on the surface, turbulenceIncrease the averaging time to 10-30 seconds
Linear deviation over the entire rangeOffset of the zero pointRecalibrate points 4 and 20mA
No response to commandsA break in the communication cableCheck the integrity of the wires, the connection of the terminals
Error E14 (overheating)Electronics temperature >85°CCheck the ventilation, solar protection of the housing
Error E7 (echo too weak)Low dielectric constant of the mediumChange antenna type or switch to GWR
Intermittent signal in GWRDeposits on the rod, mechanical damageDisassemble and clean the entire rod

Maintenance regulations

OperationPeriodicityEstimated durationQualification of the performer
Visual inspection, antenna cleaningEvery month30 minutesLocksmith of the 3rd category
Full cleaning, calibration checkEvery 3 months2 hoursElectric locksmith of the 4th category
Full service with dismantlingEvery 6 months4 hoursElectromechanic of the 5th degree
Major inspection with replacement of sealsAnnually6 hoursKVP specialist and A
Verification of measuring equipmentEvery 2 years8 hoursCertified metrologist

Directory of spare parts

Description of the partTypical specificationCategory UNITEC
O-ring sealing of the bodyNBR 70 Shore A, diameters 20-200mmSeals are industrial
Radar horn antennaFrequency 26GHz, material PTFE/steelSensors and sensors
Waveguide rodSteel 316L, diameter 6-12mm, length up to 20mComponents of KVPiA
Instrument cable2x1.5mm², shielded, temperature up to 105°CCable products
Terminal block4-pole, current 16A, voltage 600VElectrical materials
Electronic blockPower supply 24V DC, output 4-20mAElectronics of KVPiA
Connecting flangeDN50-DN200, PN16, material steel/stainless steel.Pipe fittings
Antenna protective coverIP67, material polyamide or aluminumCases and protection

A complete catalog of compatible spare parts and consumables is available in electronic catalog UNITEC-D. For specific details, please contact our technical department.

Normative links

  • DSTU EN 61508-1:2018 - Functional safety of safety systems related to electrical/electronic/programmable electronic devices
  • DSTU EN 60770-1:2016 - Transmitters for use in industrial processes
  • ISO 55000:2014 - Asset management. General overview, principles and terminology
  • DSTU OHSAS 18001:2010 - Occupational safety and health management systems
  • DSTU 2681-94 - Metrology. Terms and definitions
  • Manufacturers' instructions: Rosemount 5400/5408, Endress+Hauser Levelflex M, Vega VEGAFLEX series

Related Articles