Maintenance of spot welding machines: correction of electrodes, inspection of the transformer and service of the water cooling system

Technical analysis: Spot welding machine maintenance: electrode dressing, transformer inspection, and water cooling syst

Scope and purpose

This manual covers routine maintenance of spot welding machines from 15 to 200 kVA, including single-spot and multi-spot installations. Procedures are used for equipment with water cooling and transformer power according to DSTU EN ISO 14327.

Maintenance is performed every 500 working hours or when the quality of welded points decreases. Critical inspections of the cooling system are carried out weekly under conditions of intensive operation.

Safety measures

CAUTION: High-voltage equipment! Before starting work, it is necessary to perform the blocking/marking procedure (LOTO) in accordance with DSTU OHSAS 18001.

Mandatory personal protective equipment:

  • Dielectric gloves (class 0, up to 1000 V)
  • Safety glasses
  • Protective footwear with a dielectric sole
  • Clothes made of natural fibers

It is FORBIDDEN to perform live work. The minimum waiting time after turning off the power is 10 minutes to discharge the capacitors.

Necessary tools and materials

NameSpecificationQuantity
Socket wrenchM8-M20, torque 5-180 Nm1 set
Torque wrench10-200 Nm, accuracy ±3%1 pc
megohmmeter500-2500 V DC1 pc
MultimeterCAT III, 600 V1 pc
Manometer0-6 bar, accuracy class 1.51 pc
Contact thermometer0-100°C, ±1°C1 pc
Dresser of electrodesProcessing diameter 6-16 mm1 pc
Thickness gauge probe0.05-1.0 mm1 set
Insulating tapeClass A, up to 600 V2 rolls
Coolant liquidWater-glycol mixture, -20°C5 liters

Preliminary inspection checklist

Check elementWhat to checkAcceptance/rejection criteriaNotes
FeedingDisconnecting from the networkAcceptance: 0 V / Deviation: presence of voltageCheck all phases
ElectrodesVisual stateDeviations: cracks, deformation, wear >2 mmMeasure the diameter of the working surface
Cooling systemCoolant levelAcceptance: between MIN-MAX / Deviation: below MINCheck for leaks
Contact surfacesOxidation, pollutionDeviations: visible oxides, sootClean before assembly
Cable connectionsTighten the terminalsDeviations: backlash, darkeningCheck the tightening torque
IsolationCondition of insulating materialsDeviations: cracks, charringReplace damaged insulation

Step-by-step service procedure

Stage 1: Editing and maintenance of electrodes

  1. Dismantling the electrodes
    • Unscrew the clamping nuts with a torque of 25-30 Nm in the opposite direction
    • Carefully remove the electrodes from the holders
    • Inspect the seating surfaces for wear
    • Avoid hitting the electrodes - microcracks may form
  2. Measuring electrode wear
    • Measure the diameter of the working surface with a caliper
    • New electrode: 6.0±0.1 mm, marginal wear: 8.5 mm
    • Check the taper: the maximum allowable is 0.05 mm on a length of 10 mm
    • Measure the depth of holes from erosion: a maximum of 0.3 mm
  3. Editing the electrodes with the dresser
    • Install the electrode in the dresser cartridge
    • Set the rotation speed to 300-400 rpm
    • Tool feed: 0.1-0.15 mm/rev
    • Process until a diameter of 6.0±0.05 mm is obtained
    • Don't allow the electrode to overheat - use the SOR
  4. Processing quality control
    • Surface roughness: Ra ≤ 1.6 μm
    • Absence of burrs and microcracks
    • Perpendicularity of the end to the axis: 0.02 mm

Stage 2: Checking the transformer

  1. Visual inspection of the transformer
    • Inspect the housing for deformation and damage
    • Check the condition of the HV input insulators
    • Inspect the secondary winding for darkening
    • Check the tightening of bolt connections: 45-60 Nm
  2. Insulation resistance measurement
    • Connect a megohmmeter between the primary winding and the case
    • Test voltage: 1000 V DC
    • Minimum permissible resistance: 10 MΩ at 20°C
    • Measurement time: 60 seconds
    • At low values, dry the windings
  3. Checking the resistance of DC windings
    • Measure the resistance of the primary winding: nominal 0.15±0.02 Ohm
    • Measure the resistance of the secondary winding: nominal 0.008±0.002 Ohm
    • Measuring temperature: 20±2°C
    • The deviation from the nominal should not exceed 10%
  4. Checking the transformation ratio
    • Apply 100 V alternating voltage to the primary winding
    • Measure the voltage on the secondary winding
    • Calculate the coefficient: K = U1/U2
    • Nominal value: 50±2 (for standard transformer)

Stage 3: Servicing the water cooling system

  1. System tightness check
    • Visually inspect all connections and hoses
    • Check the tightening of the clamps: 8-12 Nm
    • Carry out a pressure test: 4.5 bar for 15 minutes
    • Permissible pressure loss: no more than 0.1 bar
  2. Changing the coolant
    • Drain the used liquid through the drain valve
    • Flush the system with distilled water
    • Pour in a new liquid: a mixture of 60% water + 40% ethylene glycol
    • System volume: 8-12 liters depending on the model
    • Do not use water with a high salt content - scale formation
  3. Cooling pump check
    • Connect pump power
    • Check performance: 15-25 l/min at 2.5 bar
    • Measure the consumed current: nominal 2.8±0.3 A
    • Check the absence of cavitation and vibrations
  4. Setting the flow regulator
    • Set the coolant flow rate: 3-5 l/min per electrode
    • System pressure: 2.0-3.5 bar
    • Temperature at the entrance: no higher than 25°C
    • Outlet temperature: no higher than 35°C

Post-service inspection checklist

TestExpected resultActualCounted/Not counted.
Transformer insulation resistance≥10 MΩ at 1000 V_____ MOm
Pressure in the cooling system2.0-3.5 bar_____ bar
Coolant consumption3-5 l/min per electrode_____ l/min
The diameter of the electrodes6.0±0.1 mm_____ mm
Tightening of electrode connections25-30 Nm_____ Nm
The temperature of the cooler at the outlet≤35°C at rated load_____ °C
Test weldingQuality point, diameter 4-6 mm_____ mm
Coolant leaksAbsent_____

Troubleshooting Guide

SymptomProbable causeCorrective actions
Weak weld pointWear of electrodes beyond the normEditing or replacement of electrodes, checking the clamping force
Sticking to electrodesContamination of the electrode surfaceElectrode cleaning, welding mode correction
Transformer overheatingInsufficient coolingChecking the flow of the coolant, cleaning the heat exchanger
Reduction of cooling pressureLeakage in the system or clogging of the filterEliminating leaks, replacing the filter, checking the pump
Unstable welding currentBad contacts in the primary circuitCleaning and tightening of contacts, checking cables
Low insulation resistanceWetting the windingsDrying the transformer at 60-80°C for 8-12 hours
Increased noise during operationLoosening of fasteners or wear of bearingsTightening of bolted connections, replacement of pump bearings
Corrosion of contact surfacesUse of low-quality coolantChanging the fluid, adding corrosion inhibitors

Recommended maintenance schedule

TaskPeriodicityEstimated durationQualification level
Visual inspection of electrodesEvery day5 minutesOperator
Checking the coolant levelEvery week10 minutesOperator
Editing of electrodesEvery 500 points or every week30 minutesMechanic
Complete maintenance of the cooling systemEvery 500 operating hours2 hoursMechanic
Checking the transformerEvery 1000 operating hours3 hoursElectrician
Replacing the coolantOnce a year or every 2000 hours1.5 hoursMechanic
Capital maintenanceEvery 8,000 operating hours8 hoursSpecialist
Calibration of regulatorsOnce a year2 hoursSpecialist

Directory of spare parts

Description of the partTypical specificationCategory UNITEC
Spot welding electrodesCuCrZr, cone 1:10, diameter 6-16 mmWelding electrodes
Sealing ringsNBR 70 Shore A, temperature -20/+120°CSealing elements
Coolant liquidEthylene glycol 40%, -25°C, 20 lTechnical fluids
Contact tipsM1 copper, silver coating 0.1 mmElectrical contacts
Cooling hosesEPDM, DN15, working pressure 6 barHydraulic hoses
Coolant filterMesh, stainless steel, 100 µmFilter elements
Insulating bushingsPolyamide PA6, dielectric strength 25 kV/mmInsulating materials
Flow sensor2-20 l/min, output 4-20 mAMeasuring devices

To select original spare parts and their analogues, use the electronic catalog UNITEC-D. Our technical specialists provide advice on compatibility and optimal selection of components for your equipment.

Reference materials

  • DSTU EN ISO 14327:2018 - Resistance welding. Procedures for determining the weldability of sheet materials up to 3 mm thick
  • DSTU EN ISO 18278-2:2017 - Resistance welding. Weldability. Part 2: Alternative procedures for the evaluation of sheet steels for spot welding
  • DSTU IEC 60974-1:2014 - Equipment for arc welding. Part 1: Power sources for welding
  • EN 50173:2007 - General requirements for electrical installations of industrial enterprises
  • ISO 14373:2015 - Resistance welding. Procedure and test for establishing welding parameters for spot welding

Related Articles